Hach APA 6000 Manual Del Usuario página 11

Ocultar thumbs Ver también para APA 6000:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

14. Push MENU and select SENSOR MENU>APA6000 #1 APA
6000>SOLUTIONS AND CLEANING>PRIMING>PRIME ENTIRE
SYSTEM NOW. Priming takes 7–15 minutes.
Note: When priming starts, a pop-up window shows the time that remains to
complete the prime cycle.
15. If bubbles are seen in the tubing after the analyzer is primed, do step
14
again to remove the bubbles (maximum of 4 times).
16. When the priming is complete, start the analyzer. Refer to
analyzer back into operation
automatically start sample measurement.
17. Let the analyzer readings stabilize (3–6 hours), then calibrate the
analyzer. Push MENU and select SENSOR
MENU>APA6000 #1 APA 6000>CALIBRATION>EXECUTE
CALIBRATION to start a calibration.
Put the
on page 5. The analyzer will
Figure 4 Rotary valve components
1 Socket head cap screws (3x)
2 Valve stator
3 Spacer ring
4 Valve rotor
5 Valve body
English 11

Publicidad

Tabla de contenido
loading

Tabla de contenido