The low-speed centrifuge can only be used indoors and placement conditions are as
follow:
● Keep a 30cm safe distance for the centrifuge, and don't put any hazardous
substances or let anybody stay within these 30cm around the machine.
● Please fix the bracket of the machine and don't swag. If you are using removable
brackets or a trolley, please ensure that there is a locking device to ensure the safe
use of the centrifuge.
● If you put the centrifuge against the wall or a corner, please make sure the distance
from the wall to the centrifuge posterior is 10cm and from the wall to its left and right
sides is 15cm so as to allow air circulation and heat elimination.
● Keep the centrifuge away from the windows to avoid heat and direct sunlight.
● The 4 supporting legs should sustain the centrifuge on the table-board evenly. Adjust
the level after placement.
● The environmental temperature should be between 5℃~40℃ and relative humidity
≤80%. Please keep the environment clean.

3.3 Locate the centrifuge at a fixed place

Don't move the centrifuge discretionarily once it is placed. Please re-confirm or
re-adjust, make the 4 supporting legs sustain it evenly on the table-board and ensure
that the mounting bracket and the table-board are fixed and don't swag if you move the
centrifuge.
3.4 Connect the power correctly
Please use a power cord separately for the centrifuge. Make sure that the cord is
well grounded and that power cords meet the safety requirements of the country or
region. The supply voltage and frequency should meet the requirements of this manual
or the specifications on the centrifuge.
Please use the power cord attached with the machine, connect it to the electric
outlet correctly and to the fixed main power supply. The power supply switch is "︱"
when open and "O" when it's switched off.