IND560xx Terminal Division 2 and Zone 2/22 Installation Manual
with a Division 1, Zone 0/1 or Zone 20/21 classification. For instance, if the
IND560xx terminal will be installed in a Division 2 area but the load cells will be
located within a Division 1 area, a load cell barrier is required. These barriers are
available through METTLER TOLEDO.
Applications involving a mixture of Division ratings or a mixture of Zone ratings
should be discussed with METTLER TOLEDO. Please consult your local METTLER
TOLEDO representative regarding these types of applications.
Replacement Parts
If a failure occurs in an IND560xx terminal that is installed in a Division 2 or Zone
2/22 area, note that only the correct parts can be used as replacement parts. For
installation on a Division 2 approved or Category 3 marked IND560xx terminal,
the following items must have the part number listed here:
3-2
DO NOT INSTALL, DISCONNECT OR PERFORM ANY SERVICE ON
THIS EQUIPMENT BEFORE POWER HAS BEEN SWITCHED OFF OR
THE AREA HAS BEEN SECURED AS NON-HAZARDOU BY
PERSONNEL AUTHORIZED TO DO SO BY THE RESPONSIBLE
PERSON ON-SITE.
Part Description
Main PCB - Analog
Main PCB – IDNet
AC Power Supply PCB
Keypad and front panel - Harsh
Keypad assembly - Panel-mount
ONLY THE COMPONENTS SPECIFIED IN THIS MANUAL CAN BE
USED IN THIS TERMINAL. ALL EQUIPMENT MUST BE INSTALLED IN
ACCORDANCE WITH THE INSTALLATION INSTRUCTIONS DETAILED
IN THIS MANUAL. INCORRECT OR SUBSTITUTE COMPONENTS
AND/OR DEVIATION FROM THESE INSTRUCTIONS CAN IMPAIR THE
INTRINSIC SAFETY OF THE TERMINAL AND COULD RESULT IN
BODILY INJURY AND/OR PROPERTY DAMAGE.
WARNING
Part Number
72198566
72198567
72226393
72214555
72214554
WARNING