SAFETY INSTRUCTIONS
• Do not expose the dental camera to liquid spray and do not store it in a humid environment (electrocution risk).
• When handling camera always take the appropriate hygiene measures and precautions in order to prevent cross
contamination risks.
• Infection control procedures must be observed when using accessories such as dental barriers in order to
prevent cross contamination risk from one patient to another.
• Install the dental camera in a clean, dry, and well-ventilated place.
• Disconnect the camera from the power supply if you are not going to use it for several days. Do not pull on the
cable.
• Never compress or nip the handpiece cable.
• Never expose the product to high vibrations.
• Do not drop the camera.
• In no way should this camera should be immersed in any liquid, nor should it be autoclaved.
• For each new patient, it is essential to use the dental barriers provided with the camera or provided as a
compliment. Before using the camera, make sure it does not have any sharp edges.
• The surface temperature in the light emission area can be just above 41°C (after several minutes of using it), so
avoid maintaining this emission area in contact with the patient's mouth.
• The SOPRO 717 First camera is a product using class II LEDs according to IEC 60825. To avoid any ocular risks, do
not stare directly at the LED's.
NOTE:
If the hygienic protection is torn while examining a patient or if the handpiece was "infected" while withdrawing
the hygienic protector, it is essential to totally disinfect the handpiece. In order to do this: please refer to the
maintenance chapter.
4
S o p r o 7 1 7 F i r s t • U s e r ' s m a n u a l • 2 0 1 0 - 0 3 - 2 6