Before reprocessing

The patient-specific accessories and consumables
must be removed from the device and, if necessary,
disassembled.
 Reprocess the reusable products in accordance
with the instructions in these instructions for use.
 Dispose of the disposable products.

Validated reprocessing procedures

Overview of the reprocessing procedures of the
components
Components
Surface disinfection with clean-
ing
Clickflow
Yes
Surface disinfection with cleaning
Surface dis-
Manufac-
Concentra-
infectant
turer
tion
Dismozon
Bode
1.6 %
plus
Chemie
Prerequisites:
The surface disinfectant has been prepared in
accordance with the manufacturer's instructions.
The manufacturer's instructions, e.g., regarding
shelf life or application conditions, are observed.
An uncontaminated, lint-free cloth soaked in sur-
face disinfectant is used for the cleaning surface
disinfection.
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
Damage to the device
Device malfunctions
Ensure that no liquid penetrates the device.
16
Cleaning
1
Wipe off obvious soiling with a disposable cloth
soaked in surface disinfectant. Dispose of the
cloth.
2
Wipe all surfaces. After that, there must no lon-
ger be any soiling visible.
Surface disinfection
3
Wipe cleaned surfaces again to visibly wet all
surfaces to be disinfected with surface disinfec-
tant.
4
Wait for the surface disinfectant contact time.
5
At the end of the contact time, moisten a new, un-
contaminated and lint-free cloth with water (at
least drinking water quality).
6
Wipe all surfaces until no remains of the surface
disinfectant, such as foam residues or streaks,
are visible.
Contact
7
Wait until the surfaces are dry.
time
8
Check the surfaces for visible damage and, if
15 min
necessary, replace the product.

Storage and transport

After reprocessing, there are no special require-
ments for storage and transport of the product. How-
ever, the following must be observed:
Store dry and free of dust
Avoid recontamination and damage during trans-
port
All further information on storage and transport in-
cluded in the accompanying documents must be ob-
served.

Other agents

Disinfectants
Use nationally approved disinfectants suitable for the
respective reprocessing process and the intended
application.
Surface disinfectant
The manufacturers of the surface disinfectants have
verified at least the following spectra of activity:
Bactericidal
Yeasticidal
Virucidal or virucidal against enveloped viruses
Follow the manufacturer's instructions for surface
disinfectants.
The following surface disinfectants were compatible
with the material at the time of testing:
Class of
Surface dis-
Manu-
active in-
infectant
facturer
gredient
Chlorine-
BruTab 6S
Brulin
releasing
Clorox Pro-
Clorox
agents
fessional Dis-
infecting
Bleach
Cleaner
Dispatch
Hospital
Cleaner Dis-
infectant
Towels with
Bleach
Klorsept 17
Medentech
Actichlor plus Ecolab USA
Instructions for use Clickflow
Listing
*
EPA
EPA
EPA
EPA

Publicidad

Tabla de contenido
loading

Tabla de contenido