Inspection

Perform inspections at regular intervals and observe
the following specifications.
.
Checks
Interval
Personnel respon-
sible
Inspection
Every
Service personnel
5 years
Performing inspection
1
Check completeness and readability of labels
2
Visual inspection of
connectors
gas type code conformity
the CS connection hose, if available
3
Functional check

Period of use

The flow regulator does not require any mainte-
nance. The life span is 10 years.

Disposing of the product

This product must be disposed of in accordance with
national regulations.

Changing the gas type cap

To change the gas type labeling to neutral color: Use
a sharp object to remove the ISO gas type cap and
replace it by the neutral gas type cap.
18

Technical data

Ambient conditions
In operation
Temperature
10 to 40 °C (50 to 104 °F)
Ambient pressure 700 to 1200 hPa
(10.2 bis 17.4 psi)
Relative humidity 0 to 95 %, no condensation
In storage and transport
Temperature
–20 to 60 °C
(–4 to 140 °F)
Ambient pressure 500 to 1200 hPa
(7.2 bis 17.4 psi)
Relative humidity 5 to 95 %, no condensation
To set the flow regulator to the existing ambient con-
ditions, see "Selected value" on page 15.
Flow settings
0.1 to 1 L/min (0; 0.1; 0.2;
0.3; 0.4; 0.5; 0.6; 0.7; 0.8;
1 L/min)
0.5 bis 5 L/min (0; 0.5; 1;
1.5; 2; 2.5; 3; 3.5; 4; 5 L/min)
1 to 16 L/min (0; 1; 2; 3; 4; 6;
8; 10; 12; 16 L/min)
1 to 32 L/min (0; 1; 5; 8; 12;
16; 20; 25; 28; 32 L/min)
Accuracy
(at 4.5 bar)
at 1 to 16 L/min
±20 %
at 1 to 32 L/min
±20 %
at 0.5 to 5 L/min
±20 %
at 0.1 to 1 L/min
±30 %
*
Supply pressure
4.5 bar
(4500 cmH
2
Weight
Central supply
version
Rail-mounted
version
Gas inlet
Central supply
plug
Rail clamp
Gas outlet
Material
Back with gas inlet
Housing
Selector wheel
Classification
as per EC Directive
93/42/EEC Annex IX
UMDNS Code
Universal Medical
Device Nomenclature
System –
Nomenclature for
medical devices
*
1 bar = 1 kPa x 100
O)
420 g (0.9 lb)
700 g (1.5 lb)
O
or AIR
2
NIST, O
or AIR
2
6 mm nozzle
Thread 9/16"
Nickel-coated brass
Powder-coated aluminum
PA 66
Class IIa
11-748