Linearity Calibration; Calibration Test; Balance Setup; Balance Set-Up Sub-Menu - OHAUS Adventurer AX124 Manual De Usuario

Ocultar thumbs Ver también para Adventurer AX124:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

EN-32
5.2.6

Linearity Calibration

Linearity calibration uses three calibration points, one at zero load and the others at specified loads.
With no load on the scale, press Linearity Calibration to begin the process.
The balance captures the zero point, and then prompts for the next weight.
Continue to follow the instructions on the display until the calibration is completed.
To cancel at any time, press Cancel.
5.2.7

Calibration Test

Use Calibration Test to compare a known calibration weight against the stored span calibration data.
With no load on the scale, press Cal Test to begin the process.
The balance captures the zero point, then prompts for the span value.
The display shows status, followed by the difference between the current calibration weight and the
stored calibration data.

5.3 Balance Setup

Enter this sub-menu to customize Balance functionality.
5.3.1

Balance Set-up sub-menu

Language

Auto Tare
Factory default settings are shown below in bold.
User Settings
Graduations
5.3.2
Language
Set the language displayed for menus and displayed
messages.
English
German
French
Spanish
Italian
Polish
Turkish
Czech
Hungarian
Filter Level
Date & Time
Approved Mode
ADVENTURER BALANCES
Auto Zero
Tracking