EN-34
5.3.8
Date & Time
Set the current Date and Time.
Change the format (if desired),
then enter the current value.
Press Save to confirm the new value.
5.3.9
Approved Mode
Use this menu to set the Legal for Trade status.
OFF = standard operation.
ON = operation complies with Legal Metrology regulations.
Note: When Approved Mode is set to ON, the menu settings are affected as follows:
Calibration Menu:
AutoCal internal Calibration is forced to ON and hidden. Internal Calibration and
Calibration Test functions are available. All other functions are hidden.
Balance Setup Menu:
Filter Level is locked at the current setting.
Auto Zero Tracking is limited to 0.5 Division and OFF. The selected setting is
locked.
Auto Tare is locked at current setting.
Graduations is forced to 1 Division and the menu item is hidden.
Communication Menu (Communication->Print Settings->Print Output):
Stable Weight Only is locked ON.
Numeric Value Only is locked OFF.
Communication Menu (Communication->Print Settings->Auto Print):
Auto print mode selections are limited to OFF, On Stability, and Interval.
Continuous is not available.
Data Maintenance Menu:
Export to USB is hidden
Import from USB is hidden
Lockout Menu:
Menu is hidden
Note: The security switch located at the rear of the balance must be in the locked
position to set Approved Mode to ON. The security switch must be in the unlocked
position to set Approved Mode to OFF. See Section 6.
Weighing application main screen with LFT turned ON.
ADVENTURER BALANCES