1.2 Combination Possibilities

This prosthetic component is part of the Ottobock modular system. As a res­
ult of its design, it may be combined with other prosthetic components of the
modular system. You can find possible restrictions in this section. Contact
the manufacturer with questions.
Designation
Liner
Socket adapter
2 Application
2.1 Indications for use
The product is to be used solely for lower limb prosthetic fittings.
2.2 Area of Application
Approved for a body weight of up to 125 kg.
2.3 Environmental Conditions
Allowable environmental conditions
Temperature range for use: -10 °C to +60°C
Allowable relative humidity 0 % to 90 %, non-condensing
Unallowable environmental conditions
Mechanical vibrations or impacts
Perspiration, urine, fresh water, salt water, acids
Dust, sand, highly hygroscopic particles (e.g. talcum)

2.4 Service Life

This prosthetic component has been tested by the manufacturer for 3 million
load cycles according to ISO 10328. Depending on the patient's activity
level, this corresponds to a service life of 3 to 5 years.
20 | Ottobock
Combination possibilities
Reference number
6Y94
4R22, 4R23, 4R37, 4R51, 4R54, 4R55, 4R73=A,
4R73=D, 4R74, 4R77, 4R95