Status Indicator Light - Hach NA5600 sc Instalación

Ocultar thumbs Ver también para NA5600 sc:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 152

2.5.1 Status indicator light

The status indicator light shows the status of the analyzer. Refer to
is above the display.
Color
Status
Green
The analyzer is in operation with no warnings, errors or reminders.
Yellow
The analyzer is in operation with active warnings or reminders.
Red
The analyzer is not in operation due to an error condition. A serious problem has occurred.
2.6 Items to collect
Collect the items that follow to install the instrument. The items that follow are supplied by the user.
In addition, collect all of the personal protective equipment appropriate to the chemicals that are
handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols.
Fasteners to
mount the
analyzer on a
wall if
applicable (4x)
Drill
Deionized water
(or sample
water)
3M KCl
electrolyte,
150 mL
Section 3 Installation
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
3.1 Installation guidelines
Install the analyzer:
• Indoors in a clean, dry, well-ventilated and temperature-controlled location.
• In a location with minimum mechanical vibrations and electronic noise.
• As close to the sample source as possible to decrease analysis delay.
• Near an open chemical drain.
• Away from direct sunlight and heat sources.
4
Use fasteners applicable to the mounting surface (¼-in. or 6 mm SAE J429-Grade 1 bolts or
stronger).
10 English
Table 5 Status indicator description
4
C A U T I O N
Table
5. The status indicator light
Level
Wire strippers
0.5M Sodium Nitrate,
500 mL
Diisopropylamine,
99%, 1 L (or
Ammonia 28%, 1 L)
Tape
measure
Wire cutters
10-mg/L
Sodium
Chloride
Standard,
500 mL
100 µm filter
for each
sample line
(optional)

Publicidad

Tabla de contenido
loading

Tabla de contenido