4 Scope of delivery

Quantity
Designation
1

Instructions for use

1
Adapter
Only for 2R41=*, 2R48 tube adapter
4
Set screw
Spare parts/accessories (not included in the scope of delivery)
Quantity
Designation
Special adhesive with hardener
Single component pack for 2R40=1 consisting of: 1
cap screw, 1 flat washer
Single component pack for 2R40=2 consisting of: 1
cap screw, 1 flat washer

5 Preparation for use

CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
► Observe the alignment and assembly instructions.
CAUTION
Improper assembly of the screw connections
Risk of injury due to breakage or loosening of the screw connections
► Clean the threads before every installation.
► Apply the specified torque values.
► Follow the instructions regarding the length of the screws and about
how to secure the screws.
5.1 Installing the foot adapter
Required tools and materials: Grinding tool, 710D4 torque wrench,
>
degreasing cleaning agent (e.g. 634A3 acetone), special adhesive and
hardener (see Page 12, table of spare parts/accessories)
1) For prosthetic feet with wooden components, roughen the connection
surface and remove the grinding dust.
12 | Ottobock
Reference number
506G3=M6x10
Reference number
636W28
636W18 and 636W19
2D6=M8
2D6=M6