Nonin 7500FO Manual Del Operador página 22

Ocultar thumbs Ver también para 7500FO:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 200
The Model 7500FO features a number of advanced options, which are
intentionally more difficult to activate. These functions are recommended
only for trained operators and require multiple button presses to prevent
accidental activation.
Function
Recall Previous
Alarm Limit
Settings
Memory
Playback
NOTE:
Alarm limits cannot be changed when the Model 7500FO is in Patient
Security mode. Patient Security mode prevents accidental changes to critical
parameters. The Model 7500FO allows users to lock and unlock alarm limits,
volume settings, and time settings.
Enter Patient
Security Mode
Exit Patient
Security Mode
18
Button
Press the Limits button while the unit is on.
"rCL" appears, indicating that previous
alarm limit settings may be recalled. To
recall the settings, press the Plus button
and select "yES." Press the Limits button
again to confirm.
Press and hold the Plus (+) button while
turning on the Model 7500FO. This
functions with the NONIN nVISION®
software. Select the Model 2500 option in
+
nVISION software (the Model 2500 option
also applies to the Model 7500FO).
To enter Patient Security mode, press and
hold the Alarm Silence button while turning
on the device.
+
To exit Patient Security mode, press and
hold the Alarm Silence and Limits buttons
while turning on the device.
+
+
Operator Functions
Instruction

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido