12
5

Formula weighing

Note: On certified scales the formula weighing functions are deactivated (factor weighing, formula weighing with
correction).
5.1
Select factor
To weigh a larger or smaller quantity of a given formula, a factor (multiplier) can be selected.
*
C.o.r.r.0.0
Factor
0. 5
fctr
*
C.o.r.r.0.0
Factor
5. 0
fctr
C.o.r.r.0.0
Factor
.
0
. . C . U . M . . I . M . M . 8 .
adv
.
0
. . C . U . M . . I . M . M . 8 .
adv
g
adv
To activate factor weighing, briefly press the «Factor» key.
The factor used for the last factor weighing (e.g. 0.50)
appears in the lower left corner of the display.
Now press the «Factor» key briefly several times until the
desired factor (0.20 ... 5.00) is displayed.
In the subsequent weighing, a component is weighed to its
target weight (e.g. 1000 g) according to the formula. How-
ever, depending on the factor selected, the actual quantity
weighed is greater or smaller by the factor.
Example 1: Weight according to formula 1000 g, factor
selected 0.50 (for half quantity). Display: 1000; actually
weighed: 500 g.
Example 2: Weight according to formula 1000 g, factor
selected 3.00 (for triple quantity). Display: 1000; actually
weighed: 3000 g.
To deactivate factor weighing: Press the «Factor» key and
hold it down.
The factor display disappears and the factor is reset to 1.