Paint Scale Panda7, Panda7/X, Panda7/X2
22
Display/Settings
/
IFACE1
IFACE2
ModE
dIALOG
dIALOG
dIALOG
dIALOG
dIALOG
2nd. d IS
PN
Print
CYCLE
Protok
HONOFF
HONOFF
HONOFF
HONOFF
HONOFF
NO
PAritY
7 EVEN
7 NO P
8 NO P
8 NO P
8 NO P
8 NO P
8 NO P
7 Odd
BAUd
300
38400
Aut. M od
AUt. S IR
AUt. S IR
AUt. S IR
AUt. S IR
AUt. S IR
NO
rESEt
Std On
EndIF1
LiSt
End
6.3
Adjusting/calibrating the scale
M.A.S.C.A.L
M.A.7.0.0.0
+
Yes
Function/Notes
Configure interface(s)
Select mode
Communication with computer (factory setting IFACE 1)
Control of auxiliary display (factory setting IFACE 2)
PM scale emulation
Communication with printer
Data printout when weight changes
Select protocol
Xon/Xoff protocol (factory setting)
No protocol
Select bits and parity
7 data bits with even parity
7 data bits with no parity
8 data bits with no parity (factory setting)
7 data bits with odd parity
Select data transmission rate
300, 600, 1200, 2400, 4800, 9600 (factory setting),
19200 and 38400
Automatic mode
Interface continuously transmits data (factory setting IFACE 1)
Automatic mode switched off (factory setting IFACE 2)
Reset IFACE setting to factory settings
Reset factory settings with «Yes», reject with «No».
Quit "IFACE1" or "IFACE2" block with «Yes»
Print out Master Mode settings with «Yes»
Quit Master Mode with «Yes». Answer "StorE?" with «Yes» to save settings or «No»
to reject them.
(not available on certified scales)
Call up Master Mode and select "CAL".
Unload the weighing pan and then press the «Yes» key to start
the calibration procedure.
kg
The scale flashes the calibration weight (corresponding to the
nominal load of the scale). If desired, the «No» key can be
used to select other calibration weights.
Place the calibration weight shown in the display on the scale
and confirm with «Yes».
Note: Calibration can be terminated at any time by pressing
+
the «On/Off» key and holding it down.
Yes
Wait until the calibration terminates successfully (the display
shows "done") and the scale returns to weighing mode.