Guidance and manufacturer's declaration – electromagnetic immunity for EQUIPMENT and SYSTEMS that are not
LIFE-SUPPORTING
The Blood pressure monitor Medel Control is intended for use in the electromagnetic environment specified below. The
customer
or the user of Blood pressure monitor Medel Control should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Conducted RF
3 V
rms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
100029_95142MEDELcontrol_userMANUAL_REV.02_22.03.2017.indd 128
Compliance
Electromagnetic environment – guidance
level
Portable and mobile RF communications equipment should be used
no closer to any part of the Blood pressure monitor Medel Control,
including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance.
3 V
rms
3 V/m
128
80 MHz to 800 MHz
800 MHz to 2.5 GHz