Recommended separation distances
Recommended separation distances between portable and mobile RF communications equipment and the
EQUIPMENT or SYSTEM – for EQUIPMENT or SYSTEM that are not LIFESUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the Blood pressure
monitor Medel Control.
The Blood pressure monitor Medel Control is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Blood pressure monitor Medel Control can help prevent elec-
tromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Blood pressure monitor Medel Control as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output
power of transmitter
(W)
0,01
0,1
1
10
100
100029_95142MEDELcontrol_userMANUAL_REV.02_22.03.2017.indd 130
Separation distance according to frequency of transmitter (m)
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.1167
0.2334
0.3689
0.7378
1.1667
2.3334
3.6893
7.3786
11.6667
23.3334