Performing A Test; Automatic Analysis; Manual Grading - Greenlee Textron GVIS 300 Manual De Instrucciones

Boroscopio de inspección de video/ dispositivo de acceso portátil
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Performing a Test

Automatic Analysis

To perform automatic grading, connect the fiber to the GVIS 400-HDP and focus the image. If the fiber needs
cleaning, do so and reconnect the fiber to the GVIS 400-HDP, focus, and press the Analyze button. The GVIS
software performs and reports the analysis by comparing defects to the IEC 61300-3-35 2009 specification and
reports Pass or Fail with the number of defects that caused a failure in each zone.
Defects that are colored in blue were found but are less than the specified size for the zone and are not
considered a failure. Defects colored red are equal to or larger than the specified size and cause a failure. See
IEC 61300-3-35 2009 specification for detailed information.
To show detailed information about a defect, mouse-over the defect so that a pop-up window appears below
the Pass/Fail box in the upper left hand corner. See Figure 19.
To change the specification that is used to determine pass or fail, choose Test Settings tab and choose the
desired specification before performing a test.

Manual Grading

Grading Obstruction Overlay
The overlay is used to judge the size of any particles or scratches in the grading zones. The zones are based on
the IEC 61300-3-35 specification.
Click the Show Grading button to display grading circles as illustrated below. The overlay can be moved by
clicking and dragging on the top bar of the overlay. To move the overlay to the home position, the lower right-
hand corner of the image, click the Home button.
Greenlee / A Textron Company
GVIS 300 • GVIS 400 • GPAD 250
Figure 19. Automatic Grading Screen
25
4455 Boeing Dr. • Rockford, IL 61109-2988 USA • 815-397-7070