9.2. MAINTENANCE
MAINTENANCE
GCE suggest to perform yearly periodic inspection by the product owner.
The inspections include leakage and functional check of the panel.
EN
LEAKAGE AND FUNCTIONAL CHECK
Perform the test according to chapter 6.2
REPAIR
Some repair activities concerning replacement of damaged or missing com-
ponents can be carried out by the owner of the product. The following com-
ponents can be replaced only:
•
labels
Contact our customer service for appropriate component number. Users
cannot exchange the plate with the CE mark and serial number by them-
selves, without GCE´s consent.
If any leakage or defect is detected, stop using it,mark that it is not al-
lowed to be used and arrange for a service repair following the instruc-
tions listed in chapter 9.3
9.3. SERVICE AND REPAIRS
The service and repair shall be carried out by a GCE authorised person or
an authorised service technician only.
Never try to disassembly the product if it is connected to the pressure
source.
Repair activities cover the replacement of the following damaged or miss-
ing components:
•
Quick couplers,
•
O-rings and sealings,
•
Gas pressure indicator,
•
Inlet and Outlet connection,
•
Hose assembly,
•
Switch knob,
•
Medi-Flow® selector,
•
Suction ejector,
•
Fixing frame.
Any product sent back to a GCE authorised person for repair shall be prop-
erly packaged. The purpose of the delivery has to be clearly specified (re-
pair, overall maintenance). For product to be repaired, a short description of
fault and any reference to a claim number would be helpful.
All labels on the equipment must be kept in a good, legible condition by
the owner and user during the entire product life time.
All seals and o-rings must be kept in dry, dark and clean environment by
the owner and the user during the entire product life time.
12/191