For the assembly, fastening accessories (F) may be provided. There are two
ways of the panel assembly – exposed or recessed assembly. Complete
kits including other accessories are available.
A - INLET CONNECTION PORT
It is intended for connection of medical gas source. The panel may be
EN
equipped with one (one gas source) or two (two sources of gas, the panel is
equipped with a cylinder source switch) inlet connection ports.
B - CYLINDER SOURCE SWITCH
By the cylinder source switch, it is possible to set the gas delivery to the
panel from source cylinder "1" or "2" or no delivery – position "0".
It is an optional part of the panel. The cylinder source switch is also avail-
able as a separate part.
C - TERMINAL UNIT–QUICK COUPLER
It is intended for connection of different medical devices such as flowme-
ters, ventilators, resuscitators, suction ejectors, etc.
It is designed according to the regional standards and is gas specific. The
quick coupler is automatically closed when a probe is disconnected and
self-opened during connection. The number of quick connectors differs fol-
lowing the panel version. Instructions for connection and disconnection of
the probes can be found in the Appendix nr. 2.
D - GAS PRESSURE INDICATOR
It is indicating the internal pressure of the medical gas in the panel. The
used scale unit bars. It is an optional part of the panel, may not be available
at some versions of the panel.
E - OUTLET CONNECTION PORT
It allows the connection of other parts of the ambulance gas supply sys-
tem via flexible hoses. At versions with source cylinder switch, the outlet is
placed downstream the switch and when the switch is in position "0" there
is no gas supplied to the outlet connection port. This is an optional part of
the panel, may not be available at some versions of the panel.
F - FASTENING ACCESSORIES
Depending on the panel version and the construction of the ambulance
wall, there may be countersunk screws M5-16 and fixing frame included
similar as shown in the picture.
G - MEDIFLOW® SELECTOR OUTLET
Requested gas flow can be adjusted by the handwheel (l/min) (G1). By
the handwheel, it is possible to regulate the gas flow – adjusting value
("0"=closed). The adjusted value of the gas flow is visible in the reading
window (G2).
Gas is delivered to the user through the flow outlet user connection (G3).
The "flow outlet" user connection is usually equipped with a connection for
hose (a hose nipple) or a threaded connection (for humidifier). This outlet
is for supplying a gas flow (l/min) at an atmospheric pressure directly to the
patient, e.g., through a cannula or a facemask.