Aesculap
Full HD 3CCD Camera system PV440
6.3
Preparations at the place of use
►
Remove any visible surgical residues to the extent possible with a
damp, lint-free cloth.
►
Transport the dry product in a sealed waste container for cleaning and
disinfection within 6 hours.
6.4
Preparation before cleaning
Camera head
►
Reprocess the product immediately after use.
►
Use suitable cleaning/disinfecting agents if the product is put away in
a wet condition.
Damage to the product due to improper handling
when reprocessing!
►
CAUTION
Separate the camera head from the camera
control unit prior to reprocessing.
►
Reprocess the camera head and the camera con-
trol unit separately.
6.5
Cleaning/disinfection
6.5.1
Product-specific safety instructions for the reprocessing pro-
cedure
Risk of electric shock and fire hazard!
►
Unplug the device before cleaning.
DANGER
►
Do not use flammable or explosive cleaning or
disinfecting solutions.
►
Ensure that no fluids will penetrate the product.
6.5.2
Validated cleaning and disinfection procedure
Validated procedure
Special features
Wipe disinfection for
–
electrical devices with-
out sterilization
■
Camera control unit
■
Manual cleaning with
immersion disinfection
■
■
Camera head
■
■
10
®
Suitable cleaning brush
20 ml disposable syringe
When cleaning products with movable hinges, ensure that these are
in an open position and, if applicable, move the joint while cleaning.
Drying phase: Use a lint-free cloth or medical compressed air
(p
=5 bar)
max
Damage to, or destruction of the product caused by
mechanical cleaning/disinfection!
►
CAUTION
Only clean and disinfect the product manually.
Damage to the product due to inappropriate clean-
ing/disinfecting agents!
►
CAUTION
Only use cleaning/disinfecting agents approved
for surface cleaning. Follow the manufacturer's
instructions for the respective cleaning/disin-
fecting agent.
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
CAUTION
►
Use cleaning and disinfecting agents according
to the manufacturer's instructions which
– Ensure that cleaning and disinfecting agents
are approved for use on anodized aluminum,
plastics (PPSU) and stainless steel.
– do not attack softeners (e.g. in silicone).
►
Observe specifications regarding concentration,
temperature and exposure time.
►
Do not exceed the maximum allowable temper-
ature of 60 °C.
Reference
Chapter Wipe disinfection for electrical devices
without sterilization
Chapter Manual cleaning/disinfecting and sub-
section:
■
Chapter Manual cleaning with immersion
disinfection