Aesculap
Full HD 3CCD Camera system PV440
4.1.5
Screwing on the endo-lens
Note
Different endo-lenses are available for the camera system.
►
Follow instructions for use of the endo-lens.
►
Screw endo-lens 1 onto camera head 5.
4.2
Settings
4.2.1
Access to main menu
►
Unplug the camera plug 6 from the camera socket 13.
►
Switch on camera control unit 7.
The following test chart appears on the monitor:
►
Press and hold Push button
The following message appears on the camera display:
4.2.2
Changing video formats
Note
If the aspect ratio is set to 4:3, the image is cut off on the left and right mar-
gins with SD video signals.
►
Select the camera frequency 50 Hz or 60 Hz (serial number 2000 and
higher): Put 50/60 Hz switch 18 in the corresponding position, accord-
ing to your requirements.
►
Press Push button
11 in the main menu to change the video format.
6
®
10 for 3 seconds.
The following formats can be set:
Camera setting
Main menu
50/60 Hz
switch
HD 1080i–
50 Hz
SD 16:9
HD 1080i–
SD 4:3
HD 720p–
50 Hz
SD 16:9
HD 720p–
SD 4:3
HD 1080i–
60 Hz
SD 16:9
HD 1080i–
SD 4:3
HD 720p–
60 Hz
SD 16:9
HD 720p–
SD 4:3
* There is always a progressive scan signal on the DVI-D output.
4.2.3
Changing the display language
►
Press Push button
14 to change the display language.
– English
– Deutsch
– Français
– Español
– Italiano
4.2.4
Exit main menu
►
Press Push button
10 or plug in camera plug 6.
Video outputs
Aspect ratio 16:9
S-Video/Video
DVI-D*
HD-SDI/
RGB
1080p 50 1080i 50
576i PAL 16:9
576i PAL 4:3
720p 50
720p 50
576i PAL 16:9
576i PAL 4:3
1080p 60 1080i 60
480i NTSC 16:9
480i NTSC 4:3
720p 60
720p 60
480i NTSC 16:9
480i NTSC 4:3