Working With The Full Hd Camera; System Set-Up; Function Checks; Safe Operation - Braun Aesculap PV440 Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 46
5.

Working with the Full HD Camera

5.1

System set-up

Position the camera tower as close as possible to the patient so that
the camera cable length of 3 m can be fully utilized.
Connect the camera plug 6 to camera socket 13 at the front panel of
camera control unit 7. Take care that the marking UP on the camera
plug 6 is pointing upwards.
5.2

Function checks

Check camera control unit 7 and the camera head 5 for external dam-
age and signs of knocks or other violent impacts.
Check that the camera cable 4 is not broken, kinked or twisted.
Press Push button
15 to power on the camera system.
The current settings of the camera appear on the monitor:
Check to make certain that the front window of endo-lens 1 is com-
pletely clean. If necessary, clean the window with isopropyl alcohol
(70 %).
If necessary, ensure that the zoom adjustment ring turns freely without
resistance.
Make certain that the focusing ring turns freely without resistance.
Align camera head 5 to an object in the room, focus the image and
check for good image quality.
To terminate the operation of the Full HD camera system press the
Push button
15.
5.3

Safe operation

Risk of injury and/or malfunction!
Always carry out a function check prior to using
WARNING
the product.
Risk of injury due to unapproved configuration!
Make certain that application parts of other
DANGER
electrical medical devices or accessories used in
this configuration for the endoscopic applica-
tion are of type BF or CF.
5.3.1

Connecting the endoscope and sterile cover

Make certain that the locking slide is opened ("Unlock" position).
Press and hold down lock.
Insert sterile endoscope with sterile cover and release lock.
To lock the endoscope against rotation, close the locking slide ("lock"
position).
Cover the camera head and cable with the sterile cover fixed to the
sterile endoscope.
Note
The endocoupler can be positioned as required. To do this, swivel endocou-
pler to the required position!
5.3.2

Performing white balance

Point endoscope to a white object (for example a sterile swab) at a dis-
tance of 2 cm.
Press Push button
12 on the control unit and keep endoscope
pointed at the white object.
- or -
Press gray Push button 2 on the camera head for longer than 2.5 sec-
onds and keep endoscope pointed at the white object.
The following message appears on the camera display and on the mon-
itor:
If the white balance is performed properly, the following message
appears on the camera display and on the monitor:
The image will be displayed in pure white. This procedure ensures the
accurate representation of all colors.
If the white balance did not perform properly, the following message
appears on the camera display and on the monitor:
In this case please repeat the white balance procedure.
5.3.3

Selecting the mode

To achieve the optimum image quality with endoscopes for different sur-
gical applications, the following modes can be selected:
Laparoscopy Standard
Laparoscopy Advanced
Arthroscopy
Hysteroscopy
Neuroendoscopy
User 1
User 2
User 3
7

Publicidad

Tabla de contenido
loading

Tabla de contenido