Technical Data - MIETHKE 75100000 Instrucciones De Uso

Ocultar thumbs Ver también para 75100000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 56
Reader Unit Set

TECHNICAL DATA

Description
Voltage range
Reader Unit
Plug-in power unit
Power consumption
Reader Unit
Plug-in power unit
Operating frequency
Bandwidth
Type of modulation
Transmitting power of the antenna
Pressure measuring range
Precision
Temperature range for pressure mea-
sur.:
Reader Unit
M.scio
Functional distance between the antenna
and M.scio
Battery life
Battery charging cycle
Battery self discharge
Flammability category of housing
Moisture protection or leak-tightness:
Reader Unit
Antenna
plug-in power unit
Durability – impact test
Durability – drop test
Weight:
Reader Unit
Antenna
Plug-in power unit
INSTRUCTIONS FOR USE | GB
Specifications and standards
6 V (DC)
100-240 V (50–60 Hz)
1.4 A (DC)
0.15-0.3A (50–60 Hz)
133 kHz
125 kHz – 135 kHz
Amplitude modulation
max. 0.8W
-66.67 hPa to +133.33 hPa
±5.32 hPa
10°C to 40°C
20°C to 39°C
10 to 30 mm
5 years minimum
at least 250 charging cycles
Remaining charge after 3 months (storage) >
70%
UL 94 HB
IP44
IP44
IP40
pursuant to IEC 60601-1: 2012 15.3.2
pursuant to IEC 60601-1: 2012 15.3.4.1
0.600 kg
0.215 kg
0.127 kg
49