Owner's Guide
B. In Beaker
For specialized item cleaning (e.g.,
prostheses): Place item into a beaker
(Fig. 11), add desired cleaning
solution (e.g., BioSonic Cement
Remover or BioSonic Ammoniated
General Purpose) according to the
products directions. Suspend the
beaker in the tank by the use of the
beaker holder and place the beaker
positioning ring around the beaker so
that the bottom 2/3 will be suspended
in the main bath.
C. Rack Cleaning
Place instruments to be cleaned in
cassette. Then place cassette in wire
rack and place the rack directly in the
tank containing BioSonic General
Purpose, Germicidal, or Enzymatic
Cleaner. Clean according to
instructions on bottle cleaner label
(Fig. 8 & 9).
D. Instrument Draining
The UC125H is equipped with a
bezel surrounding the tank. The bezel
drains condensation back into the
tank keeping the work area around
the unit dry. The bezel also acts as a
draining stand for the basket. Simply
position the corner of the basket on
two of the draining tabs (Fig. 4) at
opposite corners of the tank and
allow the solution to drain off
instruments.
IX. Accessory Use Directions
A. Full Size FingerGuard™ BioSonic
UC151XD Instrument Basket cleans
large volumes of instruments and
increases instrument processing
efficiency.
Use Directions:
Place instruments across ridges in
bottom of basket. Place basket in
tank and clean for desired time.
After cleaning rinse basket and
instruments thoroughly under
running water.
B. Three-Quarter FingerGuard: The
UC152XD Basket provides you with
the versatility of cleaning a basket
load of instruments and beaker
cleaning for prostheses (or other
small items) simultaneously.
Use Directions:
Place Basket into UC125H (Fig. 5).
Place instruments across ridges in
bottom of basket. Place basket in
tank and clean for desired time.
C. Beaker Holder
Place three UC153 Beaker Holders
on the bezel. Place positioning rings
around the beaker so that the bottom
2/3 of the beaker is suspended in the
main bath. Clean for desired time.
X. Environmental Conditions
• Intended for indoor use.
• Maximum altitude 2000 m.
• Temperature range 5°C to 40°C
(41°F to 104°F).
• Maximum relative humidity 80% for
temperatures up to 31°C (88°F),
decreasing linearly to 50% relative
humidity at 40°C (104°F).
• Main supply voltage fluctuations not
to exceed ± 10% of the nominal
voltage.
• Transient over voltages category II.
• Pollution degree 2.
• Class I equipment.
XI. Warnings
• Do not operate the unit without
sufficient solution in the tank.
• Do not use water alone as the
carrier bath. Water is not a
satisfactory coupling agent unless a
wetting agent is used. For best
results use BioSonic General
Purpose, BioSonic Germicidal
Cleaner or BioSonic Enzymatic
Ultrasonic Cleaning solutions.
• Do not place objects directly on
bottom of the tank because they
may block cavitation and cause
buildup of heat in the transducer
causing premature failure. Use only
BioSonic UC125H accessory
baskets, cassette racks or beakers
to hold items.
• Do not use flammable liquids with
BioSonic units.
• Do not restrict airflow. Provide
adequate ventilation by maintaining
at least 2 inches clearance at
the back.
• Do not use acids in the stainless
steel tank.
• Do not subject unit to shock or
impact.
• Do not immerse the unit in water.
• Keep unit covered to reduce
airborne contaminants.
• This product is intended for use only
by trained healthcare professionals.
• Use this product in accordance with
the instructions in the owner's guide.
• Unspecified or improper use of this
product may impair safety protection.
• Do not place fingers or hands in the
ultrasonic tank while in operation.
• Before cleaning items of
considerable value ultrasonically,
consult the item's manufacturer for
cleaning precautions.
• Use only the supplied power cord or
equivalent.
• The power off selection from the
keypad does not turn off main power
to the unit.
• Operation for long periods of time
will cause the cleaning solution to
get very hot. To avoid scalding, do
not place hands in the solution or
spill any onto you. Be careful when
removing an item from the tank as it
may also be hot.
• Do not disassemble the unit. There
are no user serviceable parts inside.
XII. Order Information
Cat.No
Description
Unit and accessories:
UC125H
100 Unit (100V, 50/60 Hz)
115 Unit (115V, 50/60 Hz)
230CE Unit (230V, 50/60 Hz)
230UK Unit (230V, 50/60 Hz)
240 Unit (240V, 50/60 Hz)
UC155XD Accessory Kit
3/4 FingerGuard Basket, (2) Beakers,
(2) Covers, (2) Positioning Rings and
(2) Beaker Holders
UC151XD Full Size FingerGuard Basket
UC152XD 3/4 FingerGuard Basket
UC153
One Beaker Holder
UC350
Replacement Drain Filters
UC360
Replacement Drain Hose
UC53
600 ml Beaker with Cover and
Positioning Ring
5
Qty
1
1
1
1
1
1
1
1
3
1
1