Insulation tests should only be performed on dead circuits. Check the test leads before testing.
Switch rotary to Insulation position. if meter display
Insert test leads to High、Low terminals.if the meter display Live, indicator the meter cannot measure on live circuit. please
power off Live circuit.
Press TEST.V button to select output voltage.
In insulation measuring, the screen alternate turn
on or off, the primary screen display the resistance value, and the
secondary screen display the Output voltage. Release the test button then discharges through the Meter, the meter display
DISC.
The secondary Display 0 VDC indicator when the voltage discharge finished.
Disconnect the test leads from circuit.
, please replease battery.
19
To avoid false readings, which could lead to possible electric shock or personal injury, replace the batteries as soon as
appears.
Turn the rotary switch to OFF and remove the test leads from the terminals.
1
Screw
2
Battery
20