Safety Information

This style of digital multimeter is designed and manufactured according to the safety requirements set out by the EN61010-1,
EN61010-2-030 standards for electronic test instruments . Its design and manufacture is strictly based on the provisions in
the 1000V CAT III, 600V CAT IV of EN61010-1, EN61010-2-030and the Stipulation of 2-Pollution Grade.
CAT III: It is applicable to test and measuring circuits connected to the distribution part of the building's low-voltage MAINS
installation.
CAT IV: It is applicable to test and measuring circuits connected at the source of the building's low-voltage MAINS installation.
Warning
To avoid possible electric shock or personal injury, follow these guidelines:
Use the Meter only as specified in this manual or the protection provided by the Meter might be impaired.
Do not use the Meter or test leads if they appear damaged, or if the Meter is not operating properly. If in doubt, have the
Meter serviced.
Always use the proper terminal, switch position, and range for measurements before connecting Meter to circuit under test.
Verify the Meter's operation by measuring a known voltage.
Do not apply more than the rated voltage as marked on the Meter, between the terminals or between any terminal and earth
ground.
Use caution with voltages above 30 V ac rms, 42 V ac peak, or 60 V dc. These voltages pose a shock hazard.
Replace the battery as soon as the low battery indicator (
Disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity, diodes, or
capacitance.
Do not use the Meter around explosive gas or vapor.
When using the test leads, keep your fingers behind the finger guards.
Remove test leads from the Meter before opening the Meter case or battery door.
Never operate the Meter with the cover removed or the battery door open.
Comply with local and national safety requirements when working in hazardous locations.
Use proper protective equipment, as required by local or national authorities when working in hazardous areas.
Use only the replacement fuse specified or the protection may be impaired.
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be
impaired.
) appears.
03
Accessories
Item
Test Leads
Clips
Quantity
2
2
Battery AA LR6
Manual
6
1
04