784 KFP Titrino
Result is out of limits.
Exit: The error message disappears on next start or on recalculation.
Sample size is out of limits.
Exit: The error message disappears on next start or on introduction of new
sample size.
Validation interval is expired.
Exit: The error message disappears on next start or clear counter with
&Config.Monitoring.Validation.ClearCount $G.
Service date is reached.
Exit: The error message disappears on next start or change date in
&Config.Monitoring.Service.Date.
No Oven parameters: Oven not (correctly) connected.
Exit: The error message disappears on next start. If you don't wish oven
parameters in your report, select &Mode.Parameter.Presel.Oven "no" in your
method(s).
Transmission error from Remote Box. Unknown characters.
Exit: Rectify error and switch Titrino off and on again.
Time-out error from PC keyboard (Remote Box)
Exit: Rectify error and switch Titrino off and on again.
Check Remote Box. Remote Box not (properly) connected but activated in
&Config.Periph.RemoteBox.
Exit: Rectify error and switch Titrino off and on again.