3.2.2 Description of the remote control commands
784 KFP Titrino
3.2.2.1.
Mode
Start and stop ($G, $S) or hold of the current method (3.2.2.3) with $H and
continue with $C.
$G also serves to continue after inquiries of identifications and sample size
after the start (see 3.2.2.26).
3.2.2.2.
Mode.QuickMeas
Start and stop of a measurement in the basic mode with the parameters
(measured quantity, measuring input) of the current method. Corresponds to
the <meas/hold> key.
With an ongoing measurement, the current mode can be started. This stops
the measurement automatically.
3.2.2.3.
Mode.Select
Mode.KFTQuantity
Selection of the standard mode. Mode and
the complete selection.
If a method is selected from the method memory, the nodes &Mode.Select
and &Mode.XXXQuantity are overwritten with mode and measured quantity of
the corresponding user method.
3.2.2.4.
Mode.Name
Name of the current method in the working memory. $Q sends 8 ASCII cha-
racters. Standard methods carry the name ********. The node can be set
read + write, see 3.2.2.62.
3.2.2.5.
Mode.Def.Formulas.1.Formula
Mode.Def.Formulas.1.TextRS
Mode.Def.Formulas.1.Decimal
Mode.Def.Formulas.1.Unit
Mode.Def.Formulas.1.Limits
Mode.Def.Formulas.1.LoLim
Mode.Def.Formulas.1.UpLim
Mode.Def.Formulas.1.Output
Mode.Def.Formulas.2.Formula
etc. up to .9
Entry of formulas. Rules for formula entry, see page 25.
Example: "(EP2–EP1)∗C01/C00"
In addition to the formula, a text for result output, the number of decimal plac-
es and a unit for the result output can be selected. "No unit" is selected with
the blank string.
In place of "RSX", a result name may be entered (.TextRS). This name is out-
putted in the report full, short, scalc full and scalc srt. It is used for the result
and the corresponding mean value.
3.2 Remote control commands
$G, $S, $H, $C
the measured quantity belong to
EPX, CXX, RSX, +, –, ∗, /, (, )
up to 8 ASCII characters
up to 6 ASCII characters
0...±999 999
0...±999 999
active, pulse, OFF
$G, $S
KFT
Ipol, Upol
read only
0...2...5
ON, OFF
67