3.2 Remote control commands
70
3.2.2.16. Mode.Parameter.TitrPara.StartV.Type
Mode.Parameter.TitrPara.StartV.V
Mode.Parameter.TitrPara.StartV.Factor
Mode.Parameter.TitrPara.StartV.Rate
Start volume.
If an absolute
start volume (abs.) has been selected, the volume in mL is va-
lid.
A
relative
start volume (rel.) is dispensed as a function of the sample size:
Start volume in mL = smpl size ∗ factor
The factor is valid.
The dispensing rate in mL/min applies to both cases. Max. means maximum
possible dispensing rate with the Exchange Unit in current use.
3.2.2.17. Mode.Parameter.TitrPara.Pause
Pause time in s. Is waited off after the dispensing of the start volume.
3.2.2.18. Mode.Parameter.TitrPara.ExtrT
Extraction time in s.
3.2.2.19. Mode.Parameter.TitrPara.Ipol
Mode.Parameter.TitrPara.Upol
Mode.Parameter.TitrPara.PolElectrTest
With Ipol, the inquiries for the polarization current in µA (Ipol) and
.PolElectrTest are valid.
With Upol, the inquiry for the polarization voltage in mV (Upol) is valid. Entry
in steps of 10 mV.
Besides .PolElectrTest is valid.
If the test for polarized electrodes is switched on, it is performed on chan-
geover from the inactive state to an active state (titration or conditioning).
3.2.2.20. Mode.Parameter.TitrPara.Temp
Titration temperature in °C.
3.2.2.21. Mode.Parameter.TitrPara.TDelta
Time interval in s for the entry of a measurement point in the list of measured
points.
abs., rel., OFF
0...999.99
0...±999999
0.01...150, max.
0...999999
0...999999
–127...50...+127
–1270...400...+1270
ON, OFF
–170.0...25.0...500.0
1...2...999999
784 KFP Titrino