3.2.2.52. SmplData.OFFSilo.Id1
SmplData.OFFSilo.Id2
SmplData.OFFSilo.Id3
SmplData.OFFSilo.ValSmpl
SmplData.OFFSilo.UnitSmpl
Current sample data.
The identifications Id1...Id3 can be used in formulas as sample-specific cal-
culation constants C21...C23.
If "no unit" is desired for the unit of the sample size, the blank string must be
entered.
3.2.2.53. SmplData.ONSilo.Counter.MaxLines
SmplData.ONSilo.Counter.FirstLine
SmplData.ONSilo.Counter.LastLine
Information on silo memory.
.MaxLines: Maximum possible number of silo lines.
.FirstLine: Lowest valid silo line.
.LastLine: Last occupied silo line.
3.2.2.54. SmplData.ONSilo.EditLine.1.Method
SmplData.ONSilo.EditLine.1.Id1
SmplData.ONSilo.EditLine.1.Id2
SmplData.ONSilo.EditLine.1.Id3
SmplData.ONSilo.EditLine.1.ValSmpl 6-digits, sign and dec.point
SmplData.ONSilo.EditLine.1.UnitSmpl
SmplData.ONSilo.EditLine.1.C24
SmplData.ONSilo.EditLine.1.C25
SmplData.ONSilo.EditLine.1.Mark
etc., up to .255
Contents of a silo line.
.Method: Method used to process the sample, from the method memory or
from the card.
.Id:
The identifications Id1...Id3 can also be used as sample-specific
calculation constants C21...C23 in formulas.
.UnitSmpl: If "no unit" is desired for the sample size, the blank string must be
entered.
.C24, .C25: Results which have been assigned to C24 and C25.
.Mark:
Mark of the silo line: "∗"=deleted line, "+"=line which is worked
off, "–"= line which is worked off and not valid for silo calcula-
tions (deleted), "/" last worked-off line, where recalculation can
still be done. Silo lines which have been worked off are "read on-
ly".
3.2.2.55. SmplData.ONSilo.DelLine
SmplData.ONSilo.DelLine.LineNum
Deletion of a silo line. The line # is deleted with &SmplData.ONSilo.DelLine
$G. If a formerly deleted line is edited again, it becomes valid (function "unde-
lete").
up to 8 ASCII characters
up to 8 ASCII characters
up to 8 ASCII characters
6-digits, sign and decimal point
up to 5 ASCII characters
read only
read only
read only
up to 8 ASCII characters
up to 8 ASCII characters
up to 8 ASCII characters
up to 8 ASCII characters
up to 5 ASCII characters
read only
read only
read only
1...255, OFF