5.3 Lifting the Patient
WARNING!
– DO NOT exceed the maximum weight limitation
(SWL) of 159 kg (25 stone). Maximum patient
weight is 159 kg (25 stone).
– Individuals that use the stand assist sling MUST
be able to support the majority of their own
weight, otherwise injury may occur.
– DO NOT lock the rear casters of the stand
assist lift when lifting an individual. Locking
the rear casters could cause the stand assist
lift to tip and endanger the patient and
assistants. Unlocking the rear casters during
lifting procedures allows the lift to stabilize
itself while the patient is initially lifted from a
wheelchair, bed of any stationary object.
– DO NOT move the patient if the sling is not
properly attached to the attachment points on
the stand assist lift.
– Make sure that the sling is properly attached
BEFORE lifting the patient. If any attachments
are not properly in place, correct the problem.
When the sling is elevated a few inches/cm off
the stationary surface and before moving the
patient, check again to make sure the sling and
all attachments are securely in place. If any
problem is found, lower the patient back onto
the stationary surface and correct the problem
- otherwise, injury or damage may occur.
– Adjustments for safety and comfort should be
made before moving the patient.
1078985-I
WARNING!
– Use the sling that is recommended by the
individual's doctor, nurse or medical assistant
for the comfort and safety of the individual
that is being lifted. Invacare patient slings are
made specifically for use with Invacare patient
lifts. It may in certain instances, be possible
to use other manufacturer's slings on Invacare
lifts. Contact your local Invacare office for more
information.
Detail A — Stand Assist Lift
Lifting the Patient
27