Specifications - DeVilbiss Healthcare 525 Serie Guía De Instrucciones

Ocultar thumbs Ver también para 525 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 14

SPECIFICATIONS

Catalog Number
Delivery Rate (Lower delivery rates available
for low flow applications)
Maximum Recommended Flow (@ nominal
outlet pressures of zero & 7 kPa)**
Outlet Pressure
Auxiliary Oxygen Port **
Electrical Rating
Operating Voltage Range
Oxygen Percentage
Operating Altitude
(tested at 70˚F {21˚C} only)
0-1500 M (0-4921 ft)
1500-4000 M (4921-13123 ft)
Operating Environment Range*
41˚F (5˚C) to 95˚F (35˚C), humidity range
of 10% to 95%
Power Consumption
Weight
Safe Working Load
Sound Pressure Level at 3 and 5 LPM
(ISO 80601-2-69)
Sound Power Level at 3 and 5 LPM
(ISO 80601-2-69)
Sound Level (ISO 8359:1996)
Alarm Sound Level
Dimensions
Maximum Limited Pressure
Operating System
The visible "low oxygen" indicator will
activate at the following level
Storage Conditions
Equipment Class and Type
Approval Body and Safety Standard
CE mark
EMC Compliance To
*
NOTE – The OSD performance at 41˚F (5˚C) to 95˚F (35˚C), 93% R.H. through voltage range on the 525DS verified at 670m.
**The maximum recommended flow is 3 LPM when an oxygen bottle is being filled with oxygen from the auxiliary oxygen port.
*** Use of the 515LF-607 low output flow meter package or other low output flow meter accessory will prevent the device from meeting the requirements of ISO-80601-2-69:2014 Section
201 .13 .2 .101 .
Specifications subject to change without notice.
SE-525K4
DEVILBISS 5-LITER SERIES
525DS, 525DS-Q
0 .5 to 5 LPM
5 LPM
8.5 ± 0.5 psig (58.6 ± 3.5 kPa)
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
115 V, 60 Hz, 3 .3 Amp
97-127 V~, 60 Hz
1-5 LPM=87%-96%
Across the voltage range:
No degradation of performance
Tested at nominal voltage only:
No degradation of performance
No degradation in performance across the
operating voltage range .
310 Watts Average
275 Watts @ 1.2 LPM & below
36 lbs. (16.3 Kilograms)
53 lbs. (24 Kilograms)
525DS
525DS-Q
50.9 dBA @ 3 LPM
46.7 dBA @ 3 LPM
50.7 dBA @ 5 LPM
46.7 dBA @ 5 LPM
525DS
525DS-Q
54.7 dBA @ 3 LPM
50.4 dBA @ 3 LPM
54.5 dBA @ 5 LPM
50.4 dBA @ 5 LPM
48 dBA (525DS)
46 dBA (525DS-Q)
> = 62 dBA
24.5"H x 13.5"W x 12"D
(62.2 x 34.2 x 30.4 cm)
Normal Condition: 9 PSIG (62.0 kPa),
Single Fault Condition: 27.6 PSIG (190.3 kPa)
Time Cycle / Pressure Swing
86% ± 3%
(The audible alarm will alert before O
below 82%. At less than 60%, the red "service
required" light will activate.)
-13˚F (-25˚C) to 158˚F (70˚C), humidity range
of 15% to 93% non-condensing, including
condensation
Class II Equipment Double Insulated;
Type B Applied Part, IP21
TUV
ANSI/AAMI ES60601-1:2005+A2 (R2012) +A1
IEC 60601-1-6:2010
IEC 60601-1-11:2015
ISO 80601-2-69:2014 ***
CAN/CSA-C22.2 No. 60601-1:14
CAN/CSA-C22 .2 No . 60601-1-6:11
CAN/CSA-C22 .2 No . 60601-1-11:15
CAN/CSA-C22 .2 No . 80601-2-69:16
No
EN60601-1-2
525KS, 525KS-LT
0 .5 to 5 LPM
5 LPM
8.5 ± 0.5 psig (58.6 ± 3.5 kPa)
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
220-230 V~, 50 Hz, 1 .55 Amp
230 V~, 60 Hz, 1 .9 Amp
187-253 V~, 50 Hz
195-253 V~, 60 Hz
1-5 LPM=93%±3%
Across the voltage range:
No degradation of performance
Tested at 230V/50Hz only:
No degradation of performance
No degradation in performance across the
operating voltage range .
230V / 50 Hz - 312 Watts Average
230V / 50 Hz - 296 Watts Average @
1 .2 LPM & below
230V / 60 Hz - 387 Watts Average
230V / 60 Hz - 369 Watts Average @
1 .2 LPM & below
36 lbs. (16.3 Kilograms)
53 lbs. (24 Kilograms)
525KS
47.9 dBA @ 3 LPM
49.6 dBA @ 3 LPM
47.9 dBA @ 5 LPM
49.4 dBA @ 5 LPM
525KS
51.6 dBA @ 3 LPM
53.4 dBA @ 3 LPM
51.7 dBA @ 5 LPM
53.2 dBA @ 5 LPM
40 dBA (50 Hz (525KS)
48 dBA (50 Hz (525KS-LT)
> = 62 dBA
24.5"H x 13.5"W x 12"D
(62.2 x 34.2 x 30.4 cm)
Normal Condition: 9 PSIG (62.0 kPa),
Single Fault Condition: 27.6 PSIG (190.3 kPa)
Time Cycle / Pressure Swing
86% ± 3%
drops
(The audible alarm will alert before O
2
below 82%. At less than 60%, the red "service
required" light will activate.)
-13˚F (-25˚C) to 158˚F (70˚C), humidity range
of 15% to 93% non-condensing, including
condensation
Class II Equipment Double Insulated;
Type B Applied Part, IP21
TUV approved for 50 Hz only to
IEC 60601-1:2012
IEC 60601-1-6:2010+A1
IEC 60601-1-11:2015
EN ISO 80601-2-69:2014
Yes
EN60601-1-2
525PS
0 .5 to 5 LPM
8.5 ± 0.5 psig (58.6 ± 3.5 kPa)
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
220-230 V~, 60 Hz, 1 .68 Amp
187-253 V~, 60 Hz
1-5 LPM=93%±3%
Across the voltage range:
No degradation of performance
Tested at 230V/60Hz only:
No degradation of performance
No degradation in performance across the
operating voltage range .
230V / 60 Hz - 334 Watts Average
230V / 60 Hz - 297 Watts Average @
1 .2 LPM & below
36 lbs. (16.3 Kilograms)
53 lbs. (24 Kilograms)
525KS-LT
45.4 dBA @ 3 LPM
45.3 dBA @ 5 LPM
525KS-LT
49.2 dBA @ 3 LPM
49.1 dBA @ 5 LPM
> = 62 dBA
24.5"H x 13.5"W x 12"D
(62.2 x 34.2 x 30.4 cm)
Normal Condition: 9 PSIG (62.0 kPa),
Single Fault Condition: 27.6 PSIG (190.3 kPa)
Time Cycle / Pressure Swing
86% ± 3%
drops
(The audible alarm will alert before O
2
below 82%. At less than 60%, the red "service
required" light will activate.)
-13˚F (-25˚C) to 158˚F (70˚C), humidity range
of 15% to 93% non-condensing, including
condensation
Class II Equipment Double Insulated;
Type B Applied Part, IP21
IEC 60601-1:2012
IEC 60601-1-6:2010+A1
IEC 60601-1-11:2015
EN ISO 80601-2-69:2014
EN60601-1-2
5 LPM
drops
2
TUV
Yes
EN - 11

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

525ds525ds-q525ks525ks-lt525ps

Tabla de contenido