This medical device conforms to this EN60601-1-2 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• The device needs to be installed and put into service in accordance with the information
provided in the user manual.
• The use of accessories and cables other than those specified by Manufacturer, with the
exception of cables sold by Manufacturer as replacement parts for internal components, may
result in increased emission or decreased immunity of the device.
• The medical devices should not be used adjacent to or stacked with other equipment. In
case adjacent or stacked use is necessary, the medical device should be observed to verify
normal operation in the configuration in which it will be used.
• Refer to further guidance below regarding the EMC environment in which the device
should be used.
Wireless communications equipment such as wireless home network devices, mobile phones,
cordless telephones and their base stations, walkie-talkies can affect this equipment and
should be kept at least a distance d away from the equipment. The distance d is calculated
by the MANUFACTURER from the 800 MHz to 2,5 GHz column of Table 6 of IEC 60601-1-2,
Recommended separation distance between portable and mobile
RF communications equipment and the thermometer
The thermometer are intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customers or the users of these thermometer can
help prevent electromagnetic interference by maintaining a minimum distancebetween
portable and mobile RF communications equipment (transmitters) and the thermometer
as recommended below, according to the maximum ouptut power of the communications
equipment.
Output Power of
Transmitter in Watt
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
Note: At 80MHz and 800MHz, the separation distance for the higher frequency range
applies
Note: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
Separation distance according to frequency of transmitter
in meter
150 kHz to
80 MHz
d = 1,2 √P
0.12
0.38
1.2
3.8
12
34
80 MHz to
800 MHz
d = 1,2 √P
0.12
0.38
1.2
3.8
12
800 MHz to
2,5 GHz
d = 2,3 √P
0.23
0.78
2.3
7.3
23