Instrument characteristics:
HARDWARE CHARACTERISTICS
Display
CPU
Archive memory
Calendar
Serial interface
Dosage relay
Available in connection box
Alarm relay
Available in connection box
Dosage Enabling
Power Supply
Absorption
Protection rating
Casing
Weight
Operating pressure
Maximum pressure
ENGLISH
48
Touch 7'' LCD 800 x 480 pixel
Processor 32 bit 80Mhz - 512Kb Flash – 128Kb Ram – 1Mb EEprom
Flash 512KB serial
Clock Calendar with Backup Battery.
Clock frequency 32,768 Khz
RS485
4 ON / OFF Free
(max 15A 230Vac
resistive load, max 5A 230Vac inductive load, with
interference suppression varistors).
See table 1 for configuration
1 In Exchange (Cumulative for all the Alarms)
(max 6A 230Vac
resistive load, max 1.5A 230Vac inductive load, with
interference suppression varistors)
1 Active Digital Input +5Vdc – already wired
100-240Vac 50/60 Hz
Max 12VA
IP65 (controller)
594x754x197 mm (LxHxP)
About 10 Kg
>0.5 atm with Δatm in/out >0.5atm
5 atm