Harsh Enclosure - Mettler Toledo IND560 Guia De Instalacion

Ocultar thumbs Ver también para IND560:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

4. Cut an opening in the panel or industrial enclosure per the panel cutout
dimensions shown in inches and [mm] in Figure 2-6.
Figure 2-6: Panel Cutout Dimensions
5. Place the terminal through the cutout from the front and secure with the
clamping brackets and Allen-head screws. The screws should be tightened to
a torque of 5 inch pounds (0.55 N-m).
NOTE: Once the Allen-head screws have been tightened, and the unit secured into
place, the rear cover plate of the IND560 panel mount unit may be difficult to
remove and replace for servicing. If this occurs, slight loosening of the Allen-head
screws should allow for removal and replacement of the rear cover for service
purposes.
By design, the rear cover plate can produce outward force on the (extended)
retaining brackets, increasing the overall stiffness of the enclosure structure and
providing additional assurance that the panel mount unit will remain securely in
place.

Harsh Enclosure

The harsh enclosure is made of stainless steel with a front panel angle of
approximately 38 degrees. The harsh enclosure is designed to rest on a flat
surface such as a table or desk top, or it can be mounted to a vertical surface with
optional mounting brackets.
Figure 2-5: Front Panel Gasket
IND560 Installation Manual
2-5

Publicidad

Tabla de contenido
loading

Tabla de contenido