When removing the W1 jumper for 3mV/V operation, reposition it on just one
of the pins, as shown at left in Figure 2-53. If this jumper is not correctly
positioned, the plastic cover over the Main board will not fit properly. If this
occurs, do not force the plastic cover down. Remove the cover and position
the jumper as shown.
The 2 mV/3 mV jumper settings are described in Table 2-4.
Table 2-4: Millivolt Jumper Description
Jumper
W1

Capacity Label Instructions

The regulations in some locations require that the scale capacity and increment be
shown on the front of the terminal, near the display. To satisfy this requirement, a
blue capacity label is included with the terminal that must be completed and
adhered to the front overlay.
The capacity label (shown in Figure 2-54) provides space for the Max, min, and e
information for each range or interval for which the scale is programmed. If only
one or two ranges are used, the unused portion of the label may be cut off with
scissors. Written information must be legible and a minimum of 2mm or 0.08 in.
in height. A permanent marker should be used for this information.
Figure 2-53: W1 Millivolt Jumper Settings
Setting
OPEN
Sets the connected scale's load cell setting to 3 mV/V
CLOSED
Sets the connected scale's load cell setting to 2 mV/V.
Figure 2-54: Preparing the Capacity Label
IND560 Installation Manual
Description
Cut at dotted line to remove
unused ranges/intervals
2-33