CARE AND MAINTENANCE

Proper care and maintenance of your blood pressure monitor will help
ensure long life for the product; improper use of the monitor may void the
included warranty.
Monitor Care and Maintenance
Do not use any liquids on the monitor; use a soft, dry cloth to clean the
monitor as needed
Storage outside of stated storage temperature may result in measurement
error or device malfunction; storage environment temperature is: -4°F –
140°F (-20°C – 60°C); Humidity: 10% to 93%RH (non-condensing)
Remove the batteries if the unit will not be used for an extended period
of time
Do not attempt to disassemble your monitor; disassembling your monitor
may void the manufacturers warranty
Do not subject the monitor to strong shocks; take care not to drop the
monitor
No component of this device can be maintained by the user; the circuit
diagrams, component part lists, descriptions, calibration instruction or
other information which will assist the user's appropriately qualified
technical personnel to repair those parts of equipment which are
designated repairable can be supplied
Cuff Care and Maintenance
The cuff should not be submerged or washed in a washing machine, but
may be spot cleaned with a mild-detergent as needed
Never attempt to iron or dry the cuff; if the cuff becomes wet, allow the
cuff to air dry completely before use or storage
To disinfect the cuff, wipe the inside (skin contact side) of the cuff using a
cloth lightly moistened with 70% - 90% ethyl rubbing alcohol and allow
to air dry; it is suggested to disinfect the cuff frequently when used in
clinical sittings or if the cuff is contaminated in any way.