3 Preparation

3.1 Unpack the device

3.2 Set up the device

Operating Manual 5201.6693.14 - 03.00
Remove the two protective hoods from the front and rear and inspect the device
carefully for any damage.
If the device is damaged, you should notify the responsible freight carrier immediately
and retain all items of packaging in order to support your complaint. The original
packaging is also useful for use when shipping or transporting at a later date. You
should at least keep the two protective hoods for the front and rear, to protect against
damage to the control elements and connections.
Failure to comply with any of these points may damage the equipment.
Before starting the R&S ENV216 for the first time, ensure that
an additional PE conductor is connected,
the ventilation openings are unobstructed,
no signal or operating voltages above the permissible limits are present at the
inputs,
the device's outputs are not overloaded or incorrectly connected.
Improper or negligent handling could result in death!
Note the instruction regarding protective grounding in chapter 3, section entitled
Protective grounding!
Allice Messtechnik GmbH
Preparation
6