Power Connector and Pin Out

24 Volt Extension Cables
Note: Use of an extension cable may invalidate the unit's UL Classification.
Use only copper wire for power wiring. Power extension cable shall be constructed in accordance with applicable
hospital and electrical codes. The table below shows the maximum cable length for a given wire gauge.
Cable should be constructed of SJT, SJTO, SJO, ST, SO, STO or equivalent flexible cord per US National
Difference Clause 59.1DV. Users outside of the U.S. should adhere to the applicable electrical codes for their
country.
The cable requires one each of the following connectors: Switchcraft SL403F (female) and SL413M (male). See
illustration below.
AWG
14
16
18
15
24 volt connector
Pin
Maximum Length
250 ft / 76 m
150 ft / 45 m
100 ft / 30 m
1
2
3
+ 24
GND
Shield
VDC