Caution!
Inadequate supply of oxygen possible. Keep the flow under regular observation
and adjust it again if necessary. If the device is used for a prolonged period without
interruption, the set flow rate will gradually fall as cylinder pressure drops. This can
lead to a risk to the patient.
Multistep pressure reducer (Fast)
4.3 After operation
Caution!
Never empty the oxygen cylinder completely. Always ensure that there is a certain
residual pressure in the cylinder when you return it for filling, as this prevents moist
ambient air from entering the cylinder and causing corrosion.
max.
Specified
min.
200 bar
Cylinder pressure (contents gage)
Set the multistep adjuster 4 to the desired value.
Caution!
Inadequate supply of oxygen possible.
Ensure that the multistep adjuster "engages" properly
in the desired setting on OXYWAY Fast pressure reduc-
ers. If the flow adjuster is between two adjacent set-
tings, no oxygen will flow out. This can lead to a risk to
the patient.
1. Check the remaining oxygen content on the contents
gage 1. You should change the cylinder in good time,
e.g. when the pressure drops below 50 bar, to ensure a
sufficiently long operating time.
100 bar
50 bar
EN
Operation
17