6. Functional Checks

After every change of cylinder it is essential to perform a functional check on the pressure
reducer.
If this functional check reveals any faults, you must not use the pressure reducer.
A full functional check comprises:
Visual inspection for mechanical damage;
"6.2 Checking the system for leaks" on page 22;
"6.3 Checking for leaks in the safety valve" on page 23.
We recommend you always have the following spare parts available:
Replacement seal set WM 1148;
Filter screw WM 30905

6.1 Intervals

After every change of cylinder, and at least every 6 months if not used
in the meantime
Perform a functional check.
Immediately if dirty, at least every two years
Change filter screw 11 in connecting union 2 (see "8.1 Replacing the fil-
ter screw" on page 25).
Functional Checks