6.3 Checking for leaks in the safety valve

1. Slowly open the valve on the oxygen cylinder.
2. Set the variable pressure reducers to the maximum flow.
3. Check with your finger whether oxygen is escaping at
the safety valve 10.
If oxygen is escaping, you must have the pressure re-
ducer repaired by the manufacturer (WEINMANN) or by
a trained and expert authorized dealer.