BIOTRONIK CardioMessenger Smart Manual Tecnico página 41

Ocultar thumbs Ver también para CardioMessenger Smart:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

38
Appendix
Test of
resistance
to interfe-
rence
Electro-
static
discharge
(ESD)
according
to IEC
61000-4-2
Fast
transient
electric in-
terference
(bursts)
according
to IEC
61000-4-4
Surge
according
to IEC
61000-4-5
Test level
Com-
according
pliance
to IEC
60601-1-2
±6 kV
±8 kV
contact
contact
discharge
dis-
± 8 kV air
charge
discharge
± 15 kV
air dis-
charge
±2 kV for
Same
power
as test
supply
level
lines
±1 kV for
input and
output
lines
±1kV
Same
push-pull
as test
voltage
level
kV
common-
mode
voltage
Guidelines for the elec-
tromagnetic environ-
ment
Floors should be made
of wood or cement, or
have ceramic tiles.
When the floor consists
of a synthetic material,
the relative humidity
must be at least 30 %.
The quality of the supply
voltage should corre-
spond to that in a typical
business or hospital.