NOTE: Choose between the following
cleaning methods.
In case the practice is not equipped with
a thermodisinfector, the cleaning process
can begin with the manual pre-washing
(paragraph 06.1.1) and proceed with the
sterilization (Chapter 09).
In case the practice is equipped with a
thermodisinfector the pre-washing can be
performed manually (paragraph 06.1.1) or
step is always the sterilization (Chapter
09).
06.1
MANUAL/AUTOMATED CYCLE
06.1.1
MANUAL PRE-WASHING AND CLEANING WITH
ENZYMATIC DETERGENT
06
CLEANING OF THE ACCESSORIES
have been validated by the medical
device manufacturer as ABLE to reprocess
a medical device for being reused. It is
up to the process responsible to assure
that the repeated processes are actually
performed by means of equipment,
materials and personnel structured as
repeated processes in order to obtain the
desired result. Generally this requires the
validation and the systematic monitoring
of the process. Similarly, in case the
person in charge of the process deviates
from the supplied instructions, he shall
properly evaluate any action and their
consequencies.
Prepare an enzymatic* detergent solution
1
with pH 6-9, as per the instructions of the
manufacturer;
dispose of the enzymatic detergent
solution, do not recycle.
HANDPIECE WITH CORD
Clean the surface of the handpiece and
2
of its cord with a clean, soft cloth with
detergent solution (pH 6-9) and, if you
wish to disinfect use a non-aggressive
disinfectant solution with neutral pH (pH7),
following the instructions provided by the
manufacturer of the solution;
CAUTION: The following instructions
CAUTION: Once used, correctly
EN
13