02
CLEAN/FLUSH FUNCTION
The CLEAN/FLUSH function (according to the device used) enables to run a cleaning cycle
on the irrigation circuit of the handpiece used during the treatment, by following the steps
described in this chapter.
EN
CAUTION: Clean
The CLEAN/FLUSH function must be used
after every treatment, before starting the
cleaning and sterilization procedures.
irrigation bag and from the peristaltic
pump tube or from the tube of the bone
grafting kit, if used.
following the norms in force locally on the
disposal of hospital wastes;
Prepare an enzymatic detergent solution
with pH 6-9, as per the instructions of the
manufacturer;
Immerge the end part of the peristaltic
pump tube or the tube of the bone
grafting kit, if used, in a container
containing the enzymatic detergent
solution prepared;
4
ush function.
1
2
3
CAUTION: If the tubes are not cleaned,
the crystallization of the salts may seriously
damage the device.
CAUTION: The handpiece and the cord
cannot be separated.