Calibrating An Oxygen Instrument - MSA ALTAIR Pro H2S Manual De Operación

Detector de un sólo gas
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Calibrating an Oxygen Instrument

False oxygen alarms can occur due to changes in barometric
pressure (altitude changes) or extreme changes in ambient
temperature.
The ALTAIR Pro Detector is equipped with a feature to allow
calibration at pressure and/or temperature of use.
Be sure that the instrument is in known fresh air before performing
a calibration.
To enter the calibration mode, make sure you are in fresh,
uncontaminated air.
1. In normal operating mode, hold the TEST button for two seconds.
See FIGURE 4-2 for more details.
2. When the "TEST" "GAS" "?" displays, press and hold the TEST
button to enter calibration.
• "TEST" "CAL" displays
After three seconds, "FAS" "?" displays asking the user if a
fresh air setup/calibration is desired.
3. Press the TEST button at the "FAS" "?" screen to perform a
calibration at 20.8% O 2 .
NOTE: This procedure must be performed in fresh,
uncontaminated air. Do not breathe on the sensor while
performing this function.
If the successfully calibrates:
"OK" displays
If the sensor does not successfully calibrate:
"ERR" displays.
4. Wait five seconds.
The instrument returns to normal operating mode.
5. If "ERR" displays after calibration, the current settings did not
change. Immediately check that:
The instrument is in fresh, uncontaminated air during the
calibration process
No one breathes on the sensor during calibration.
• Repeat steps 1 through 6, as necessary.
6. The display must read "OK"; if "ERR" remains, remove the
instrument from service.
7. Perform a bump test to confirm operation and activate the " ".
4-5

Hide quick links:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Altair pro o2Altair pro co

Tabla de contenido