3.
Point the syringe tip upward, place the syringe into the bracket
such that the attachment sleeve is above the bracket.
4.
Hold the attachment sleeve in the bracket and screw the syringe
onto the valve from below.
With the narrow side facing down, carefully press the attachment
sleeve into the bracket.
5.
Withdraw the piston from the syringe and place it onto the screw in
the dosage arm. Hold the syringe piston and turn the attachment
screw on the dosage arm into the thread in the piston.
6.
Fix the syringe cover with six screws.
W A R N I N G
Risk of injury to eyes, skin and respiratory
During the first days of system use, check the screw joint between the syringe and
valve to avoid leaks and the potential escape of reagents.
When working with chemicals and/or solvents, always observe the relevant
accident prevention regulations and wear appropriate personal protective clothing
for eye, face, hand, body and/or respiratory protection.
English
8
Figure 2 Installation of the burette
Installation of the titration stand
The titration stand consists of a holder onto which the electrode, titration
cell and magnetic stirrer are attached. The titration is performed in the
titration cell.
1.
Place the titration stand on the right side (syringe side) of the
titrator (observe the routing of the cable).
2.
Place a magnetic stirring bar into a titration cell.
Place a titration cell over the magnetic stirrer.
3.
Place the rubber ring on the stand rod to set the minimum height of
the electrode holder.