Initial Use of the Device

Always check all parts and accessories of the device immediately after receiving
the shipment. Inspect for external damage as well. The manufacturer considers
only replacement claims that have been immediately submitted or reported to a
sales representative or an authorized service company.
Install the device on a level surface and in a dry environment. The ambient
temperature has to be between 10°C and 40°C; the relative humidity has to be
between 30% and 75%.
The device is not explosion-proof. Do not use in an area where flammable
anesthetic gases are present.
4.1

Mains Connection

Check to make sure that the available mains voltage matches the data listed on
the label attached to the back of the device (type plate). Incorrect voltage can
cause errors and malfunctions and may destroy the equipment.
Make sure the mains power supply is installed in compliance with DIN VDE or
international rules and regulations. The power cord must be plugged into a
properly installed safety wall plug (see DIN VDE 0107). Read the device label
located in rear of device (type plate) to determine the operating voltage of the
device.
To avoid risk of electric shock, this equipment must only be connected to a
supply mains with protective earth.
The power connection has to be grounded. Use the original power cable to
establish a connection between the mains wall socket and the non-heating
device plug located in the rear of the device.
Only use a certified (UL listed), removable mains connection line, type SJT,
minimal 18 AWG, 3 leads. The plugs have to comply with NEMA 5-15 and/or IEC
320/CEE22. Grounding will only be reliable if the equipment is connected to a
corresponding hospital grade socket.
Integrate the device into the potential equalization system as specified by local
safety rules and regulations.
Make sure that neither the mains cable nor the equipotential cable touch the
scale system. This could lead to an error in balancing. The device may only be
operated with angular equipotential MC plugs.
Initial Use of the Device
Input control
Setup
DANGER
WARNING
Protective contact
WARNING
Only for U.S. operators
Potential equalization
DANGER