E. PREPARING THE MOTOR HANDPIECE FOR USE
1. Connecting the Power Cord
Securely insert the AC power cord into the
charging stand. Securely insert the AC
power cord into a wall receptacle. (Fig. 3)
2. Charging the Battery
The X–SMART
™
into the charging stand, the light turns yellow indicating the unit is charging. When
the battery is fully charged, the light will turn green.
PRECAUTION
• Connect to AC 100-240V power only. Otherwise, malfunction will occur.
• The standard charging time is about 1 1/2 hours, but it varies depending on use, condition of
battery, temperature, and if the battery is new or old.
• Notice: Older battery shortens both the charging and operating time.
3. Connecting and Disconnecting the Contra-Angle Attachment
• The contra-angle attachment can be connected to
the motor handpiece at 6 adjustable head positions.
Align the positioning pins of the contra-angle
attachment with the positioning slots of the motor
handpiece and insert the head until they click.
• When removing the contra-angle attachment, pull it
straight out. (Fig. 4)
PRECAUTION
• Before attaching and detaching the contra-angle attachment, turn the power off.
• Check that the contra-angle attachment is securely assembled to the motor handpiece.
• Before first use complete the calibration per Section E.5.
12
is delivered with a rechargeable battery. Put the handpiece
Easy
Fig. 3
Fig. 4