TORQUE
• The torque limit values can be adjusted.
• Torque can be found in the top box.
• If the torque limit is exceeded, the file will stop and "
box on the display.
Display
SPEED
• The rotation speed can be adjusted.
Display
AUTO REVERSE
The auto reverse mode can be selected. (See Auto Reverse Function, section F.5.)
The default mode for auto reverse is ON. The last position used for auto reverse is
what will be the new default until the user manually changes it. Only the mode that is
currently in use will be displayed.
Display
The present torque limit value is displayed. The
torque setting can be adjusted from 0.6 N•cm to
1.0 N•cm in increments of 0.2 N•cm, and from
1.0 N•cm to 4.0 N•cm in increments of 0.5 N•cm.
The present rotation speed is displayed in the
middle box on the motor handpiece. The speed
setting can be adjusted from 100 to 800 rpm in
increments of 50 rpm.
ON
If, during operation, the load reaches the
torque limit, the unit auto reverses. When the
load is removed after auto reverse rotation,
it returns to normal forward rotation again.
If during operation, the load reaches the
ON OFF
torque limit, the unit auto reverses. When the
load is removed after auto reverse rotation, it
stops.
Auto reverse rotation is not activated.
OFF
" is shown in the torque
15